Command Reference
R&S
®
ZNC
438
User Manual 1173.9557.02 ─ 13
Manual operation:
See
CALCulate<Chn>:RIPPle:CLEar
Resets the limit check results for the ripple test.
Suffix:
<Chn>
.
Channel number
Usage:
Event
Manual operation:
See
6.3.1.13
CALCulate:SMOothing...
The
CALCulate:SMOothing...
commands provide the settings for trace smoothing.
..........................................................................438
.............................................................................438
CALCulate<Chn>:SMOothing:APERture
<SmoothAperture>
Defines how many measurement points are averaged to smooth the trace.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<SmoothAperture>
Smoothing aperture. A smoothing aperture of n % means that the
smoothing interval for each sweep point i with stimulus value x
i
is
equal to [x
i
- span*n/200, x
i
+ span*n/200], and that the result of i
is replaced by the arithmetic mean value of all measurement points
in this interval.
Range:
0.05% to 100%.
*RST:
1
Default unit: %
Example:
*RST; :CALC:SMO ON
Activate smoothing for the default trace.
CALC:SMO:APER 0.5
Reduce the smoothing aperture to 0.5 %.
Manual operation:
See
CALCulate<Chn>:SMOothing[:STATe]
<Boolean>
Enables or disables smoothing for trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference