GUI Reference
R&S
®
ZNC
232
User Manual 1173.9557.02 ─ 13
All Channels Continuous / on Hold
Selects a common sweep control mode for all channels of the active recall set.
●
"All Channels Continuous:" The R&S ZNC continuously repeats the sweeps in all
channels.
●
"All Channels on Hold:" The R&S ZNC performs single sweeps, according to the
channel-specific number of "Sweeps".
In R&S ZVAB compatibility mode the buttons are unavailable, however, you can use the
"Restart Manager" dialog instead.
Remote command:
4.4.2.6
Restart Manager (Dialog)
The "Restart Manager" dialog defines whether the active sweep mode ("Continuous" or
"Single") and the "Number of Sweeps" are valid for all channels in the active recall set or
for the active channel only.
This dialog is relevant / accessible in the "ZVR" or "ZVABT" compatibility modes only.
Related information
Refer to
chapter 4.4.2.5, "Sweep > Sweep Control"
Access:
CHANNEL > SWEEP > Sweep Control > Restart Manager...
Sweep All Channels
Apply the sweep control settings to all channels in the active recall set. The number of
sweeps in a "Single" sweep sequence is equal to the selected number of "Sweeps" times
the number of channels. The sequence starts with the first sweep in channel no. 1.
Tip:
In remote control, it is possible to retrieve the results acquired in any of the sweeps
within a single sweep group.
Remote command:
ALL
Channel Settings