User Manual 1173.9557.02 ─ 13
Calibration Unit
Displays the connected calibration units. The R&S ZNC auto-detects all calibration units
which are connected to one of its USB ports. If several cal units are connected, one of
them must be selected for characterization (active cal unit).
Remote command:
Authentication / Set Password
Set a password to protect the characterization dialog and the "Start Characterization"
wizard from unauthorized access and operation. "Set Password" opens a dialog to enter
the password and activate password protection at the next time the "Characterize Cal
Unit" dialog is opened. Enter an empty string (no password) to deactivate password pro-
tection.
Tip:
A password also blocks a switchover of the active characterization during calibration;
Remote command:
Characterization Data
Displays all characterizations which are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications.
Channel Settings