Command Reference
R&S
®
ZNC
380
User Manual 1173.9557.02 ─ 13
Parameters:
<TraceName>
String parameter containing the trace name
<Format>
FDATa | SDATa | MDATa | NCData
Data format; see
Data format identifiers used in the CALCu-
.
Example:
See
Data format identifiers used in the CALCulate:DATA... com-
Usage:
Query only
6.3.1.3
CALCulate:DLINe...
The
CALCulate:DLINe...
commands control the horizontal line used to mark and
retrieve response values (display line).
.................................................................................................380
......................................................................................380
CALCulate<Chn>:DLINe
<Position>
Defines the position (response value) of the horizontal line.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Position>
See list of parameters below.
Range:
-200 dB to +200 dB
*RST:
0 dB
Default unit: dBm
Example:
*RST; :CALC:DLIN 10
Define the position of the horizontal line in the default dB Mag dia-
gram at +10 dB.
CALC:DLIN:STAT ON
Display the defined horizontal line.
Manual operation:
See
CALCulate<Chn>:DLINe:STATe
<Boolean>
Switches the horizontal line on or off.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<Boolean>
ON | OFF - horizontal line on or off
*RST:
OFF
Example:
See
SCPI Command Reference