Command Reference
R&S
®
ZNC
614
User Manual 1173.9557.02 ─ 13
[SENSe<Ch>:]SWEep:DWELl
<MeasDelay>
Defines the "Meas. Delay" before each partial measurement or the first partial measure-
ment (depending on whether "All Partial Meas'ments" or "First Partial Meas'ment" is
selected). Setting a delay disables the automatic calculation of the (minimum) sweep
time; see
.
Suffix:
<Ch>
.
Channel number
Parameters:
<MeasDelay>
Measurement delay before each partial measurement. Changing
the delay leaves the number of points unchanged but has an
impact on the duration of the sweep (
Range:
0 s to 13680 s
*RST:
0 s
Default unit: s
Example:
*RST
Reset the instrument, activating a frequency sweep with the S-
parameter S
21
as a measurement result for channel and trace no.
1.
SWEep:TIME?
Query total sweep time.
SWEep:DWEL 1
SWEep:DWEL:IPOint ALL
Set a delay of 1 s for each partial measurement.
SWE:TIME?
Query total sweep time. The time is extended by the delay times
the total number of sweep points (one partial measurement per
sweep point required).
Manual operation:
See
[SENSe<Ch>:]SWEep:DWELl:IPOint
<InsertionPoints>
Defines whether the measurement delay (previously defined via
) is inserted before all partial measurements or before the first partial
measurement only.
Suffix:
<Ch>
.
Channel number
Parameters:
<InsertionPoints>
ALL | FIRSt
Insertion before all or before the first partial measurement
*RST:
ALL
Example:
See
Manual operation:
See
SCPI Command Reference