Command Reference
R&S
®
ZNC
410
User Manual 1173.9557.02 ─ 13
CALCulate<Chn>:MARKer<Mk>:MODE
<Mode>
Sets marker no.
<Mk>
to continuous or discrete mode. The marker doesn't have to be
created before (
ON
), the mode can be
assigned in advance.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters:
<Mode>
CONTinuous | DISCrete
CONTinuous - marker can be positioned on any point of the trace,
and its response values are obtained by interpolation.
DISCrete - marker can be set to discrete sweep points only.
*RST:
CONT
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK:MODE DISC; :CALC:MARK2:MODE CONT
Create marker 1 in discrete mode and marker 2 in continuous
mode.
CALC:MARK ON; MARK2 ON
Display the two markers. Due to the different modes the horizontal
positions can be different.
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:NAME
<MarkerName>
Defines a name for marker no.
<Mk>
. The marker doesn't have to be created before (
ON
), the name can be assigned in advance.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters:
<MarkerName>
Marker name (string parameter)
*RST:
'M1' for marker no. 1 etc.
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK:NAME '&$% 1'; :CALC:MARK ON
Create marker 1 named "&$% 1" and display the marker .
CALC:MARK:REF ON
CALC:MARK:REF:NAME 'Reference'
Display the reference marker and rename it "Reference".
Manual operation:
See
SCPI Command Reference