Command Reference
R&S
®
ZNC
592
User Manual 1173.9557.02 ─ 13
6.3.14.10
[SENSe:]PORT...
The
[SENSe:]PORT...
commands define the reference impedances at the physical
ports.
[SENSe<Ch>:]PORT<PhyPt>:ZREFerence
<RealPart>[, <ImaginaryPart>]
Specifies the complex reference impedance for the physical port numbered
<PhyPt>
.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Physical port number
Parameters:
<RealPart>
Real part of the port impedance.
Range:
1 mΩ to 10 MΩ
*RST:
Default reference impedance of the connector family
assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
<ImaginaryPart>
Imaginary part of the port impedance. This part may be omitted to
define a real impedance.
Range:
- 10 MΩ to 10 MΩ
*RST:
Default reference impedance of the connector family
assigned to the port (real impedance, e.g. 50 Ω).
Default unit: Ohm
Example:
PORT2:ZREF 52, 2
Specify a complex reference impedance of 52 Ω + j * 2 Ω for the
(physical) port no. 2.
CALC:TRAN:IMP:RNOR PWAV
Select renormalization of port impedances according to the power
waves theory.
Manual operation:
See
"Renormalization according to Theory of"
6.3.14.11
[SENSe:]POWer...
The
[SENSe:]POWer...
commands configure the automatic/adaptive gain control
(AGC) of the R&S ZNC.
See also
..................................................................................593
.....................................................................593
.......................................................................594
...........................................................................594
.........................................................595
...............................................................595
SCPI Command Reference