GUI Reference
R&S
®
ZNC
113
User Manual 1173.9557.02 ─ 13
Some of the softkeys in the "S-Params" tab open dialogs.
●
see
chapter 4.2.1.2, "S-Parameter Wizard"
●
see
chapter 4.2.1.3, "Balanced Ports Dialog"
S-Parameter
Selects a scattering parameter (S-parameter) as measured quantity for the active trace.
For a channel setup with n logical ports, one of n
2
S-parameters can be selected.
The S-parameters are the basic measured quantities of a network analyzer. They
describe how the DUT modifies a signal that is transmitted or reflected in forward or
reverse direction. For background information see
Single-ended (unbalanced) S-parameters are referred to as S<out>< in>, where <out>
and <in> denote the output and input logical port numbers, respectively.
In presence of balanced ports, S-parameters are defined in the form
S<m_out><m_in><out><in>, where output mode <m_out> and input mode <m_in> can
be one of
●
d (differential, balanced)
●
c (common, balanced)
●
s (single-ended, unbalanced)
Remote command:
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22" ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "S11" | "S12" |
"S21" | "S22" ...
S11 / S12 / S21 / S22
Selects one of the four elements of the standard 2-port scattering matrix (S-parameters)
as a measured quantity for the active trace.
Trace Settings