GUI Reference
R&S
®
ZNC
235
User Manual 1173.9557.02 ─ 13
Fig. 4-1: Calibration Unit Wizard: Ports
Ports
Selects the test port(s) to be calibrated.
Remote command:
The port parameters in many calibration commands define the calibrated port(s).
Cal Unit
Displays the connected calibration units. The R&S ZNC auto-detects all calibration units
which are connected to one of its USB ports. If several cal units are connected, one of
them must be selected for calibration (active cal unit).
The calibration unit R&S ZV-Z51 is suited for R&S ZNC network analyzers. A warning is
displayed if the current sweep range of the network analyzer exceeds the characterized
frequency range of the calibration unit.
See also
chapter 3.5.5.1, "Connecting the Calibration Unit"
Remote command:
Characterization
Displays all characterizations which are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications. To account for a modified test setup (e.g. the connection of addi-
tional adapters to the calibration unit), you can generate modified sets of characterization
data using the cal unit characterization wizard; see
chapter 4.4.3.10, "Characterize Cal
, on page 260. By default, the R&S ZNC uses the last generated cal unit
characterization.