The circuit model for the load consists of capacitance C which is connected in parallel to
an inductance L and a resistance R, both connected in series.
●
"R" is the constant resistive contribution. It is possible to select a special value
("Open" for
∞
Ω so that the inductance coefficients are irrelevant, "Short" for 0 Ω,
Match for the reference impedance of the current connector type) or set any resist-
ance R.
●
The fringing capacitance C and the residual inductance L are both assumed to be
frequency-dependent and approximated by the first four terms of the Taylor series
around f = 0 Hz.
See also description of the load parameters for the different standard types in
ter 3.5.2.1, "Calibration Standard Types"