Command Reference
R&S
®
ZNC
426
User Manual 1173.9557.02 ─ 13
Usage:
Event
Manual operation:
See
CALCulate<Ch>:PARameter:DELete:SGRoup
Deletes a group of logical ports (S-parameter group), previously defined via
.
Suffix:
<Ch>
.
Channel number.
<Ch>
may be used to reference a previously
defined channel. If
<Ch>
does not exist, it is generated with default
channel settings.
Example:
See
Usage:
Event
CALCulate<Ch>:PARameter:MEASure
<TraceName>[, <Result>]
Assigns a measurement result to an
existing
trace. The query returns the result assigned
to the specified trace (no second parameter; see example).
Note:
To create a new trace and at the same time assign the attributes, use
. To display the trace, create a diagram (
ON
) and assign the trace to this diagram (
); see example below.
Traces must be selected to become active traces; see
defined traces. You can open the "Trace Manager" dialog to obtain an overview of all
channels and traces, including the traces that are not displayed.
Suffix:
<Ch>
.
Channel number of an existing channel containing the referenced
trace.
Parameters:
<TraceName>
Trace name, string variable, e.g. 'Trc4'. See "Rules for trace
names" in
must be unique across all channels and diagrams.
<Result>
Measurement parameter (string variable); see
A query of a wave quantity
'xy'
returns
'xyD<n><Detector>'
, where
<n>
numbers the source
(drive) port, and
<Detector>
denotes the detector setting (
SAM
for a "Normal" (sample),
AVG
for an "AVG Real Imag",
AMP
for an
AVG Mag Phase
detector). A query of a ratio
'x/y'
returns
'xD<n>/yD<m><Detector>'
, where
<n>
and
<m>
number the
source ports
SCPI Command Reference