Command Reference
R&S
®
ZNC
581
User Manual 1173.9557.02 ─ 13
<PhyPt>
Calibrated port number. This suffix is ignored because the port
number is specified in the parameter list.
Setting parameters:
<Wave>
AWAVe | BWAVe | B1 | B2 | B3 | B4
AWAVe
–
calibration of reference waves a
1
, a
2
, .... In manual con-
trol, the reference receiver calibration is included in the source
power calibration of each port.
BWAVe
–
calibration of received wave b
1
, b
2
, .... The port number
<CalPort>
, the used
<SourceType>
, and the
<SourcePort>
number must be specified in addition.
Alternative: Parameters B1 and B2.
B1 | B2 ...
–
direct wave and cal port setting for received waves
b
1
, b
2
... The parameters B1 and B2 are ZVR-compatible. No addi-
tional parameters need to be specified. The source for B1 is Port
2 and vice versa.
<CalPort>
Calibrated port number
Range:
1 to port number of the analyzer
<SourceType>
PORT | GENerator
PORT - internal source at port
<CalPort>
, to be fed to port
<SourcePort>
using an external through connection. If
<CalPort> = <SourcePort>
, an Open or Short standard is
required.
GENerator - not supported.
<SourcePort>
Number of the port for the internal source or generator.
Range:
1 to port number of the analyzer/number of external
generators.
Example:
*RST; CORR:POW:ACQ AWAVe,1,PORT,1
Perform a receiver power calibration of the wave a
1
using port 1
as a source port, assuming that the transmitted source power a
1
is correct. No external cabling is needed.
CORR:POW:AWAV?
Check whether the calibration is applied (the response is
1
).
CALC:PAR:MEAS 'TRC1', 'A1'
Select a
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no.
<Mk>
in the center of the sweep range and
query the measurement value. The calibrated power of the refer-
ence wave a
1
is approx. equal to the default source power value.
See also example for
Usage:
Setting only
Manual operation:
See
SCPI Command Reference