Command Reference
R&S
®
ZNC
700
User Manual 1173.9557.02 ─ 13
POWer:S<Ptout><Ptin>
S-parameter with output and input port number of the
DUT, e.g. S11, S
21
.
POWer:RATio A<Ptout> | B<Ptin>,
A<Ptout> | B<Ptin>
Output:
A<Ptout>/B<Ptin>
Ratio, e.g. B2, A1 for b
2
/a
1
drive Port 1
POWer:A<Ptout>
Wave quantity with stimulus port number of the ana-
lyzer, e.g. a
1
.
POWer:B<Ptin>
Wave quantity with receive port number of the ana-
lyzer, e.g. b
2
.
POWer:Z<Ptout><Ptin>
Matched-circuit impedance (converted Z-parameter)
with output and input port number of the DUT, e.g.
Z
11
, Z
21
.
POWer:Y<Ptout><Ptin>
Matched-circuit admittance (converted Y-parameter)
with output and input port number of the DUT, e.g.
Y
11
, Y
21
.
POWer:KFACtor
POWer:MUFactor<Lev>
Stability factor K
Stability factors
μ
1
or
μ
2
Note:
The mnemonics POWer: and VOLTage: are not used in output strings.
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect[:ACQuire]
<Sensor>
Initiates a source power calibration for the source port
<PhyPt>
using an external power
meter no. 1 or 2. To initiate a source power calibration for arbitrary power meters, use
the alternative commands listed in the program example below.
Note:
The command cannot be used unless a power meter is connected via GPIB bus,
USB or LAN interface and configured in the "External Power Meters" dialog.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Setting parameters:
<Sensor>
ASENsor | BSENsor
ASENsor
–
external power meter Pmtr 1.
BSENsor
–
external power meter Pmtr 2.
*RST:
ASENsor
Example:
SOUR:POW3:CORR:COLL BSEN
Perform a source power calibration for port 3 using power meter
no. 2.
SOUR:POW:CORR:PMET:ID 2
Select power meter no. 2.
SOUR:POW:CORR:ACQ PORT,3
Perform a source power calibration for port 3 using the previously
selected power meter no. 2.
R&S ZVR/ZVAB Compatible Commands