GUI Reference
R&S
®
ZNC
265
User Manual 1173.9557.02 ─ 13
Efficient power calibration procedure
●
For standard applications, open the "Start Cal" tab and select "Scalar Power Cal >
Power Cal..." to perform the necessary calibration sweeps with default power cali-
bration settings. You do not need any of the buttons in the "Pwr Cal Settings" tab.
●
Select "Cal Power ..." if you use an amplifier between the source port and the DUT.
●
Select "Transmission Coefficients ..." if you want to modify the power calibration pro-
cedure.
Background information
Refer to
chapter 3.5.6, "Scalar Power Calibration"
Access:
CHANNEL > CAL key or Alt + Shift + P
The "Calibration" buttons open the following dialogs:
●
Cal Power...:
See
chapter 4.4.3.13, "Modify Cal Power Dialog"
●
Transmission Coefficients...:
See
chapter 4.4.3.14, "Power Meter Transmission
●
Power Meters...:
See
chapter 3.7.5, "External Power Meters"
Switch Off All Other Sources
Ensures that the power at all sources except the calibrated source is switched off during
the calibration. This is advisable especially if the measurement involves a combination
of different signals.
Remote command:
Channel Settings