Command Reference
R&S
®
ZNC
584
User Manual 1173.9557.02 ─ 13
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the receiver power calibration for
the received waves b
<PhyPt>
.
*RST:
OFF
Example:
*RST; CORR:POW:ACQ BWAVe,1,PORT,2
Perform a receiver power calibration of the wave b1 using port 2
as a source port, assuming that the source power a
2
is correct. A
through connection from port 2 to port 1 is needed.
CORR:POW?
Check whether the calibration is applied (the response is
1
).
CALC:PAR:MEAS 'TRC1', 'B1D2'
Select b
1
as a measured quantity for the default trace.
CALC:MARK ON; MARK:Y?; :SOUR:POW?
Create marker no.
<Mk>
in the center of the sweep range and
query the measurement value. The calibrated power of the
received wave b
1
is approx. equal to the default source power
value.
See also example for
Manual operation:
See
[SENSe<Ch>:]CORRection:SSTate?
Returns the system error correction state label of the active trace in channel
<Chn>
. The
response is a string variable containing the calibration state label in the trace list ('Cal',
'Cai', 'Cal Off' ..; see
chapter 3.5.4, "Calibration Labels"
Suffix:
<Ch>
.
Channel number used to identify the active trace.
Example:
See
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection[:STATe]
<Boolean>
Enables or disables the system error correction for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the correction.
*RST:
ON
SCPI Command Reference