GUI Reference
R&S
®
ZNC
164
User Manual 1173.9557.02 ─ 13
Time Gate
Enables or disables the time gate for the time domain and frequency domain traces.
"Gat" is displayed in the trace list while the time gate is active.
Remote command:
Axis Pair
"Start" and "Stop" or "Center" and "Span" define the size of the time gate. The analyzer
generates a warning if the selected time span exceeds the unambiguous range which is
given by Δt = 1/Δf, where Δf is the spacing between two consecutive frequency points.
Simply reduce the time span until the warning disappears.
Remote command:
Show Range Limits
Displays or hides two red lines indicating the start and stop of the time gate in the diagram.
Remote command:
Bandpass / Notch
The filter type defines what happens to the data in the specific time region.
●
A "Band Pass" filter passes all information in the specified time region and rejects
everything else.
Trace Settings