GUI Reference
R&S
®
ZNC
249
User Manual 1173.9557.02 ─ 13
After successful measurement receiver calibration, the b<n> trace should be close to the
a<m> trace. Due to the previous power calibration, both traces should be close to the cal
power.
Remote command:
Port Overview
Shows all source ports together with the possible power calibrations. Either a source
power calibration (flatness calibration, "Power") or a measurement receiver calibration
("Meas. Receiver") can be performed at each analyzer port P1 ... PN.
Select a "Power" or "Meas. Receiver" symbol to open the "Power Cal" dialog and perform
a calibration sweep. A green checkmark indicates that the calibration data has been
acquired successfully.
See also
Remote command:
n/a, see
Apply
Is enabled as soon as a new set of power calibration data has been acquired for either
a "Power" source or a "Meas. Receiver". The button applies all available source and/or
measurement receiver calibrations to the active channel, aborts the verification sweeps,
and closes the port overview section.
The power calibration state is indicated in the trace list, see
, on page 89. Use the functions in the "Use Cal" panel to activate,
deactivate, or store power calibrations.
Remote command:
Power Cal
–
Power (Dialog)
The "Power Cal
–
Power" dialog guides you through the power calibration for a particular
source port (flatness and reference receiver calibration). Measurement receiver calibra-
tion is described in
Background information
Refer to
chapter 3.5.6.1, "Source Power Calibration"
Access:
CHANNEL > CAL > Start Cal > Power Cal... > Power
Channel Settings