GUI Reference
R&S
®
ZNC
240
User Manual 1173.9557.02 ─ 13
●
the calibration may be invalid
If auto-detection fails because of a high attenuation in the signal path, you can either
enter the port assignment manually or connect matching port numbers and select "Default
Port Assignment".
Remote command:
Apply/Cancel
Apply the calculated system error correction to the active channel (or to all channels in
the active recall set, if all channels were calibrated).
The Apply button is enabled as soon as calibration sweeps have been successfully per-
formed for all required port assignments.
Remote command:
Extended cal unit settings:
4.4.3.3
Calibration Presetting Wizard
The "Calibration Presetting" wizard guides you through the setup and execution of a
manual system error correction.
The wizard proceeds through the following steps:
1.
Ports and Type:
Select the ports to be calibrated and the calibration type to be used.
2.
Connectors and Cal Kits:
Select the connector type and gender for all ports to be
calibrated. If necessary, load or modify a calibration kit.
3.
Calibration:
Acquire measurement data for the required ports or port pairs and the
required standards. Finally, decide whether or not to apply the resulting system error
correction.
Channel Settings