GUI Reference
R&S
®
ZNC
165
User Manual 1173.9557.02 ─ 13
●
A "Notch" filter rejects all information in the specified time region and passes every-
thing else.
Remote command:
Shape
Selects a gate shape which the R&S ZNC uses to filter the trace in the time domain. The
drop-down list visualizes how the time gate will affect a constant function after transfor-
mation back into the frequency domain. The selected window is applied to the active
trace. The two red vertical lines represent the "Start" and "Stop" values defining the size
of the time gate.
The R&S ZNC always uses a "Steepest Edges (Rectangle)" window to calculate the time-
gated frequency domain trace, see
"Time-Gated Frequency Domain Trace"
See also
Remote command:
Side Lobe Level
Defines the side lobe suppression for an "Arbitrary Gate Shape (Dolph-Chebychev)" gate.
The entered value is the ratio of the power of the central lobe to the power of the first side
lobe in dB.
Remote command:
4.2.4.10
Trace Config > Trace Statistics
Evaluates statistical and phase information of the entire trace or of a specific evaluation
range and calculates the x-dB compression point.
Access:
TRACE > SCALE key or Alt + Shift + D
Trace Settings