Command Reference
R&S
®
ZNC
413
User Manual 1173.9557.02 ─ 13
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value.
Parameters:
<MarkerValue>
Stimulus value of the reference marker.
Range:
-9.9E+11 Hz to +9.9E+11 Hz (for frequency sweeps);
-999 dBm to +999 dBm (for power sweeps); 0 s to
127500 s (for time and CW mode sweeps)
*RST:
0 Hz
Default unit: Hz
Example:
Suppose that the active recall set contains an active trace no. 1
and that the sweep range for a frequency sweep starts at 1 GHz.
CALC:MARK:REF ON
Create the reference marker and display it in the center of the
sweep range.
CALC:MARK:REF:X 1GHz
Set the reference marker to the beginning of the sweep range.
CALC:MARK:REF:Y?
Query the measurement value at the reference marker position.
Manual operation:
See
"Mkr Stimulus / On / Delta Mode"
CALCulate<Chn>:MARKer<Mk>:REFerence:Y?
Returns the response (in Cartesian diagrams: y-axis) value of the reference marker. The
reference marker must be created before using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value.
Example:
See
Usage:
Query only
Manual operation:
See
CALCulate<Chn>:MARKer:SEARch:BFILter:RESult[:STATe]
<Boolean>
Shows or hides the bandfilter search results in the diagram area.
Suffix:
<Chn>
.
Channel number used to identify the active trace.