Command Reference
R&S
®
ZNC
412
User Manual 1173.9557.02 ─ 13
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value.
Parameters:
<Boolean>
ON | OFF - creates or removes the marker.
*RST:
OFF
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK:REF ON; :CALC:MARK ON
Create the reference marker and marker 1 and assign them to
trace no. 1. The default position of both markers is the center of
the sweep range.
Manual operation:
See
"Mkr Stimulus / On / Delta Mode"
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE
<Mode>
Sets the reference to normal or fixed mode. The marker must be created before using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value.
Parameters:
<Mode>
NORMal | FIXed
NORMal - response value changes according to the measurement
result.
FIXed - marker keeps its current response value.
*RST:
NORMal
Example:
CALC:MARK:REF ON; :CALC:MARK:REF:TYPE FIX
Create the reference markerand display it in the center of the
sweep range as a fixed marker.
CALC:MARK:REF:X 1GHz
Shift the marker horizontally. The response value remains fixed.
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:REFerence:X
<MarkerValue>
Defines the stimulus (in Cartesian diagrams: x-axis) value of the reference marker, which
can (but doesn't have to) be displayed using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference