Command Reference
R&S
®
ZNC
564
User Manual 1173.9557.02 ─ 13
Setting parameters:
<CalName>
Name of the calibration (string parameter) defined together with
the calibration type (
ALL
- the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of
*RST:
-
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard as
calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the mea-
surement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the active
channel, then delete the data.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:FIXTure[:ACQuire]
<StandardType>,
<TestPort1>, <TestPort2>...
Starts a fixture compensation sweep in order to acquire measurement data for a test
fixture that has its inner conductor terminated with the selected standards.
Suffix:
<Ch>
.
Channel number of the corrected channel
Setting parameters:
<StandardType>
OPEN | SHORt
Terminating standard type: open or short.
<TestPort1>
<TestPort2>
Test port numbers. For a fixture compensation, n arbitrary (not
necessarily consecutive) port numbers must be specified.
*RST:
-
SCPI Command Reference