Command Reference
R&S
®
ZNC
656
User Manual 1173.9557.02 ─ 13
Example:
See
Usage:
Event
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:CATalog?
Queries the numbers of all configured external power meters. The response is a string
containing a comma-separated list of power meter numbers.
Suffix:
<Pmtr>
.
Number of the configured power meter. This suffix is ignored; the
command affects all power meters.
Example:
SYST:COMM:RDEV:PMET1:DEF 'USB Power Meter 1',
'NRP-Z55', 'usb', '100045'
SYST:COMM:RDEV:PMET2:DEF 'USB Power Meter 2',
'NRP-Z55', 'usb', '100046'
Configure two R&S NRP power meters as external power meter
no. 1 and 2, assigning names and serial numbers.
SYST:COMM:RDEV:PMET1:CAT?
Query the power meter numbers. The response is '1, 2'.
SYST:COMM:RDEV:PMET1:COUN?
Query the number of configured power meters. The response is
2.
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<Pmtr>:CONFigure:AUTO[:STATe]
<Boolean>
Enables or disables "Auto Config NRP-Zxx". If the function is enabled, the analyzer auto-
matically configures all power meters detected at any of the USB ports as Pmtr 1,
Pmtr2, ...
Suffix:
<Pmtr>
.
Number of the configured power meter. This suffix is ignored; the
command affects all power meters. Power meters can be assigned
several times so that the number of configured power meters is
practically unlimited.
Parameters:
<Boolean>
ON | OFF - enable or disable "Auto Config NRP-Zxx".
Example:
See
Manual operation:
See
SCPI Command Reference