TRIGger<Ch>[:SEQuence]:MULTiple:SOURce
<Sequence>[, <TrigSource>]
Selects the source of the trigger events that the analyzer uses to start a measurement
sequence in multiple trigger mode.
Suffix:
<Ch>
.
Channel number
Parameters:
<Sequence>
SWEep | SEGMent | POINt | PPOint
Triggered measurement sequence,
PPOint
denotes "partial mea-
.
<TrigSource>
IMMediate | EXT1 | EXT2 | E1A2 | E1O2 | MANual
Trigger source for the triggered measurement sequence:
IMMediate - free run measurement (untriggered).
EXT1, EXT2 - external trigger 1, 2.
E1A2, E1O2 - external trigger 1 and 2, 1 or 2.
MANual - trigger event generated by pressing the "Manual Trig-
ger" softkey.
Example:
TRIG:MULT:SOUR SWE, EXT1; SOUR POIN, EXT2
Select external trigger 1 as a trigger source for the entire sweep,
external trigger 2 as a trigger source for each point.
TRIG:MULT:SLOP1 SWE, POS; SLOP2 POIN, POS
Trigger on the rising edges of the external trigger signals 1 and 2.
TRIG:MULT:HOLD POIN, 1ms
Define a trigger delay of 1 ms before each sweep point.
Manual operation:
See
"Source / Slope / Level / Delay"
TRIGger<Ch>[:SEQuence]:SLOPe
<Slope>
Qualifies whether the trigger event occurs on the rising or on the falling edge or on the
beginning of the high / low level periods of the external TTL trigger signal.
Suffix:
<Ch>
.
Channel number
Parameters:
<Slope>
POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
Manual operation: