Command Reference
R&S
®
ZNC
554
User Manual 1173.9557.02 ─ 13
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<Characterization>
Name and (possibly) directory of the characterization (cal kit file)
to be used for the automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file stored
in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path denotes a spe-
cific cal kit file stored in the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path denotes a specific
cal kit file stored in an arbitrary directory.
<TestPort>
<TestPort>
Test port numbers. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified. The
analyzer automatically detects the calibration unit ports connected
to each analyzer port.
Example:
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port calibration at test ports 1, 2, and 4
using the calibration unit's default calibration kit file and arbitrary
test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AUTO:CKIT
<Characterization>
Generates a characterization (cal kit file) with the specified name containing the cal kit
data of the active calibration unit (
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
).
The cal kit file can be stored in the calibration unit or written to a directory on the analyzer.
Note
that this command can only be executed if the number if cal unit ports is less or
equal to the number of test ports. Furthermore the command assumes the "canonical"
assignment of cal unit ports to test ports: cal unit port 1 assigned to test port 1, cal unit
port 2 assigned to test port 2 etc. The new command
allows for flexible assignments.
Setting parameters:
<Characterization>
Name and (possibly) directory of the created characterization / cal
kit file (string parameter, extension *.calkit): If no path and directory
is specified, the file is stored in the calibration unit (see also
[SENSe<Ch>:]CORRection:COLLect:AUTO
). The factory cal-
ibration data on the unit is not overwritten. If a path and directory
is specified, the file is stored in the directory on the analyzer's hard
disk. The default directory (
MMEMory:CDIRectory
) is not used.
Example:
CORR:COLL:AUTO:CKIT 'AutoCalChar.calkit'
Generate a cal kit file
AutoCalChar.calkit
for the active cali-
bration unit and store it in the calibration unit.
SCPI Command Reference