Command Reference
R&S
®
ZNC
548
User Manual 1173.9557.02 ─ 13
Example:
MMEM:LOAD:CKIT 'C:\Users\Public
\Documents\Rohde-Schwarz
\Vna\Calibration\Kits\New Kit 1
(123456).calkit'
Load the previously created cal kit file
New Kit 1 (123456).calkit
from the default cal kit directory.
CORR:CKIT:LSEL 'N 50 Ohm', 'New Kit 1',
'123456'
Assign the imported kit to the N 50 Ω connector type, assuming
that the cal kit name stored in
New Kit 1 (123456).calkit
reads
New Kit 1
and that its label reads
123456
.
[SENSe:]CORRection:CKIT:STANdard:LCATalog?
<CalKitName>, <KitLabel>
Returns a list of all standards in a given calibration kit.
Parameters:
<CalKitName>
to query cal kit names and labels.
<KitLabel>
Name of the calibration kit label.
Example:
See
Usage:
Query only
[SENSe:]CORRection:CKIT:<StandardType>:WLABel
<ConnType>, <CalKitName>,
<CalKitLabel>, <StandardLabel>, <MinFreq>, <MaxFreq>, <ElLength>, <Loss>,
<Z0>, <C0>, <C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>, [OPEN | SHORt |
MATCh, <Resistance>, <Port1>, <Port2>]
Defines the parameters of a non-ideal 1 port or 2-port calibration standard
<StandardType>
. A particular physical standard can be selected by specifying the
name and label of the calibration kit and the standard's serial number. Depending on the
standard type, only a subset of the parameters may be used; see
.
Note:
If the specified cal kit does not exist, it is created with the specified calibration
standard.
SCPI Command Reference