GUI Reference
R&S
®
ZNC
122
User Manual 1173.9557.02 ─ 13
More Ratios
Opens a dialog to select arbitrary ratios of wave quantities, e.g. for different source ports
or higher port numbers.
4.2.1.5
More Ratios (Dialog)
The "More Ratios" dialog provides arbitrary ratios with arbitrary source ports as measured
quantities. All ratios can be calculated with different detector settings.
Background information
Refer to the following sections:
●
chapter 3.3.4.1, "Wave Quantities"
●
Access:
TRACE > MEAS > Ratios > More Ratios...
The notation for ratios follows the usual scheme of the vector network analyzer:
●
The a-waves are the outgoing/transmitted waves at the analyzer's test ports.
●
The b-waves are the incoming/measured waves.
●
The source port for the stimulus signal must be specified in addition.
●
The port number range covers all test ports of the analyzer.
Numerator
Selects the type (left pull-down list) and the port number assignment (right pull-down list)
of the wave that forms the numerator of the ratio.
Remote command:
"<Trace_Name>", "B2/A1" | ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "B2/A1" | ...
Trace Settings