Command Reference
R&S
®
ZNC
480
User Manual 1173.9557.02 ─ 13
Parameters:
<RelFontSize>
Relative font size
Range:
80 % to 170 %
*RST:
100 %
Default unit: percent
Example:
*RST; :DISP:RFS 80
Use smaller fonts to gain more space for the traces in the diagram.
Manual operation:
See
DISPlay[:WINDow<Wnd>]:CATalog?
Returns the numbers and names of all diagrams in the current recall set.
The response is a string containing a comma-separated list of diagram area numbers
and names, see example below. If all diagram areas have been deleted, the response is
an empty string ("").
Suffix:
<Wnd>
.
Number of a diagram. This suffix is ignored; the command returns
a list of all diagrams.
Example:
*RST; :DISP:WIND2:STAT ON
Create diagram no. 2.
DISP:WIND2:NAME 'S11 Test Diagram'
Assign a name to the new diagram.
DISP:CAT?
Query all diagrams and their names. As a default diagram no. 1 is
created upon *RST, the response is
''1,1,2,S11 Test
Diagram'
. The first diagram is not named; its default name is
equal to the diagram number.
CALC:PAR:SDEF 'Win2_Tr1', 'S11'
Create a trace named Win2_Tr1 to measure the input reflection
coefficient S
11
.
DISP:WIND2:TRAC9:FEED 'Win2_Tr1'
Display the generated trace in diagram area no. 2, assigning the
trace number 9 to it.
DISP:WIND2:TRAC:CAT?
Query all traces in diagram area no. 2. The response is
'9,Win2_Tr1'
.
Usage:
Query only
Manual operation:
See
DISPlay[:WINDow<Wnd>]:MAXimize
<Boolean>
Maximizes all diagram areas in the active recall set or restores the previous display con-
figuration.
SCPI Command Reference