GUI Reference
R&S
®
ZNC
128
User Manual 1173.9557.02 ─ 13
Tip:
Use the "Y- Z-Params" tab to measure Z-parameters including the transfer param-
eters. Use the Smith chart to obtain an alternative, graphical representation of the con-
verted impedances in a reflection measurement.
Remote command:
"<Trace_Name>", "Z-S11" | "Z-S12"
| "Z-S21" | "Z-S22"
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "Z-S11" | "Z-S12"
| "Z-S21" | "Z-S22"
Balanced Ports...
Opens a dialog to define a balanced port configuration.
See
chapter 4.2.1.3, "Balanced Ports Dialog"
4.2.1.9
Meas > Y <-- Sij
Selects converted admittances as measured quantities. The admittances are calculated
from the measured S-parameters.
Background information
Refer to the following sections:
●
chapter 3.3.3, "Admittance Parameters"
●
chapter 3.3.3.1, "Converted Admittances"
Access:
TRACE > MEAS key or Alt + Shift + A
Y <-
Selects a converted admittance parameter as a measured quantity for the active trace.
For an n-port vector network analyzer, the pull-down list provides the full set of n
2
admit-
tance parameters.
Trace Settings