GUI Reference
R&S
®
ZNC
133
User Manual 1173.9557.02 ─ 13
Balanced Ports...
Opens a dialog to define a balanced port configuration.
See
chapter 4.2.1.3, "Balanced Ports Dialog"
4.2.1.12
Meas > Stability
Selects one of the three two port stability factors K, μ
1
or μ
2
as measured quantities. A
typical application of stability factors is to assess the stability of an amplifier. Stability
factors cannot be calculated in balanced port configurations.
Background information
Refer to
chapter 3.3.6, "Stability Factors"
Access:
TRACE > MEAS key or Alt + Shift + A
Stability
Selects a stability parameter as a measured quantity for the active trace. The stability
factor calculation is based on 2-port reflection
and
transmission S-parameters so that the
input and output port numbers must be different. The pull-down list contains all possible
physical (single-ended) port combinations. For an analyzer with n ports, provides n * (n
–
1) stability parameters.
Stability parameters are expressed as "K
<out><in>
", "μ1
<out><in>
", and "μ2
<out><in>
", where
<out> and <in> denote the logical output and input port numbers of the DUT.
Remote command:
"<Trace_Name>", "KFAC21" |
"MUF121" | "MUF221" | ...
Create new trace and select trace name and measurement parameter:
"<Trace_Name>", "KFAC21" |
"MUF121" | "MUF221" | ...
Trace Settings