Command Reference
R&S
®
ZNC
429
User Manual 1173.9557.02 ─ 13
'MUF221' | 'MUF212' | ...
Stability factor 2 (for unbalanced ports only)
'Pmtr1D1' | 'Pmtr2D2'
Power sensor measurement using a power meter 'Pmtr<no>' and analyzer
source port 'D1' or 'D2'
*) Selecting a parameter Y...<n><m> or Z...<n><m> sets the range of port numbers to
be considered for the Y and Z-parameter measurement to <n>:<m>.
CALCulate<Ch>:PARameter:SELect
<TraceName>
Selects an existing trace as the active trace of the channel. All trace commands without
explicit reference to the trace name act on the active trace (e.g.
).
CALCulate<Ch>:PARameter:SELect
is also necessary if the active trace
of a channel has been deleted.
Suffix:
<Ch>
.
Channel number.
Parameters:
<TraceName>
Trace name, e.g. 'Trc4'. See "Rules for trace names" in
Example:
CALC4:PAR:SDEF 'Ch4Tr1', 'S11'
Create channel 4 and a trace named
Ch4Tr1
to measure the input
reflection coefficient S
11
. The trace is the active trace in channel
4.
CALC4:PAR:SDEF 'Ch4Tr2', 'S22'
Create another trace named
Ch4Tr2
to measure the output reflec-
tion coefficient S
22
. Again this new trace becomes the active trace
in channel 4.
CALC4:PAR:SEL 'Ch4Tr1'
Select the first trace
Ch4Tr1
as the active trace.
CALC4:FORM MLIN
Calculate the magnitude of S
11
and display it in a linearly scaled
Cartesian diagram.
6.3.1.11
CALCulate:PHOLd...
The
CALCulate:PHOLd...
commands control the max hold (peak hold) function.
CALCulate<Chn>:PHOLd
<HoldFunc>
Enables, disables, or restarts the max hold and the min hold functions.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
SCPI Command Reference