Command Reference
R&S
®
ZNC
385
User Manual 1173.9557.02 ─ 13
Suffix:
<Chn>
.
Channel number used to identify the active trace.
Parameters:
<TimeGate>
RECT | HAMMing | HANNing | BOHMan | DCHebyshev
RECT - steepest edges (rectangle)
HANN - normal gate (Hann)
HAMMing - steep edges (Hamming)
BOHMan - minimum flatness (Bohman)
DCHebyshev - arbitrary gate shape (Dolph-Chebychev)
*RST:
HANN
Example:
See
Manual operation:
See
CALCulate<Chn>:FILTer[:GATE]:TIME[:TYPE]
<TimeGateFilter>
Selects the time gate filter type, defining what occurs to the data in the specific time
region.
Suffix:
<Chn>
.
Channel number used to identify the active trace-
Parameters:
<TimeGateFilter>
BPASs | NOTCh
BPASs - band pass filter: Pass all information in specified time
region and reject everything else.
NOTCh - notch filter: Reject all information in specified time region
and pass everything else.
*RST:
BPASs
Example:
*RST; :CALC:FILT:TIME:STAT ON
Reset the instrument and enable the time gate.
CALC:FILT:TIME NOTCh
Select a notch filter in order to reject unwanted pulses.
Manual operation:
See
6.3.1.5
CALCulate:FORMat...
The
CALCulate:FORMat...
commands determine the post-processing of the mea-
sured data in order to obtain various display formats.
CALCulate<Chn>:FORMat
<Type>
Defines how the measured result at any sweep point is post-processed and presented in
the graphical display.
SCPI Command Reference