Command Reference
R&S
®
ZNC
542
User Manual 1173.9557.02 ─ 13
Example:
See
Usage:
Query only
[SENSe:]CORRection:CKIT:<StandardType>
<ConnType>, <CalKitName>,
<StandardLabel>, <MinFreq>, <MaxFreq>, <ElLength>, <Loss>, <Z0>, <C0>,
<C1>, <C2>, <C3>, <L0>, <L1>, <L2>, <L3>, [OPEN | SHORt | MATCh,
<Resistance>, <Port1>, <Port2>]
Defines the parameters of a non-ideal 1 port or 2-port calibration standard
<StandardType>
. A particular physical standard can be selected by specifying the
name of the calibration kit and its serial number. Depending on the standard type, only a
subset of the parameters may be used; see
.
Note:
If the specified cal kit does not exist, it is created with the specified calibration
standard.
Parameters:
<StandardType>
Standard type. For reflection standards, the first character denotes
the gender, e.g.:
FOPen, MOPen:
Open (f) or Open (m) standard.
The following reflection standards are supported:
MOPen,
FOPen, MSHort, FSHort, OSHort, MOSHort, FOSHort,
MMTCh, FFTCh, MREFLect, FREFLect
For transmission standards, the first two characters denotes the
genders on both ends, e.g.:
FFSNetwork, MFSNetwork, MMSNetwork:
Symm. network
(ff), symm. network (mf) or symm. network (mm) standard.
For a complete list of standard types refer to
Parameter list
String parameters to specify the configured standard
(
'<ConnType>', '<CalkitName>', '<Standard>'
) and
numeric parameters defining its properties. See
Note:
The following transmission standards have only 8 parame-
ters (until <Z0>):
MMTHrough, MFTHrough, FFTHrough,
MMLine, MFLine, FFLine
The following reflection and transmission standards have only 5
parameters (until <MaxFreq>):
MSMatch, FSMatch,
MMATten, MFATten, FFATten
*RST:
n/a
Example:
CORR:CKIT:FOP 'N 50 Ohm','New Kit 1', '',
0,4000000000,0,0,50,99,-2.3,0.22,0,0,0,0,0
Define the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
CORR:CKIT:FOP? 'N 50 Ohm'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in the active cal kit.
CORR:CKIT:FOP? 'N 50 Ohm','New Kit 1'
Query the properties of the Open (f) standard for a N 50 Ω con-
nector type in cal kit "New Kit 1".
SCPI Command Reference