Command Reference
R&S
®
ZNC
543
User Manual 1173.9557.02 ─ 13
Manual operation:
See
"Add / Copy / Delete / Standards..."
The parameters in the
[SENSe<Ch>:]CORRection:CKIT:<StandardType>
,
[SENSe<Ch>:]CORRection:CKIT:<StandardType>:WLABels
, and
[SENSe<Ch>:]CORRection:CKIT:<ConnectorType>:<StandardType>
com-
mands have the following meaning:
Table 6-5: Parameter list
Parameter
Meaning
Comment/Unit
'<ConnectorName>'
Name of the connector type. Use SENSe1:CORRection:CON-
Nection:CATalog? to query connector names.
String parameter
'<CalKitName>'
Name of the calibration kit. Use
[SENSe<Ch>:]CORRection:CKIT:CATalog?
to query cal
kit names.
String parameter
'<CalKitLabel>'
Label (e.g. the serial number) of the calibration kit;
for
...WLABel
command only
String parameter
'<StandardLabel>'
Label (e.g. the serial number) of the standard
String parameter
<MinFreq>
Min. frequency for which the circuit model is valid
Default unit is Hz
<MaxFreq>
Max. frequency for which the circuit model is valid
Default unit is Hz
<ElLength>
Electrical length (offset parameter) of the standard
Default unit is m
<Loss>
Loss (offset parameter) of the standard
To be specified without unit (implicit unit is
dB)
<Z0>
Reference impedance (no unit)
To be specified without unit (implicit unit is
Ω)
<C0> | <L0>
Polynomial coefficients C0 or L0 for the fringing capacitance or
residual inductance of the standard (load parameter)
To be specified without unit (implicit unit is
fF or pH)
<C1> | <L1>
Polynomial coefficients C1 or L1 for the fringing capacitance or
residual inductance of the standard (load parameter)
To be specified without unit (implicit unit is
fF /GHz or pH /GHz)
<C2> | <L2>
Polynomial coefficients C2 or L2 for the fringing capacitance or
residual inductance of the standard (load parameter)
To be specified without unit (implicit unit is
fF/GHz
2
or pH/GHz
2
)
<C3> | <L3>
Polynomial coefficients C3 or L3 for the fringing capacitance or
residual inductance of the standard (load parameter)
To be specified without unit (implicit unit is
fF/GHz
3
or pH/GHz
3
)
OPEN | SHORt |
MATCh
Simplified modelling as an open or short or match standard.
The load circuit model generally consists of capacitance C
which is connected in parallel to an inductance L and a resist-
ance R, both connected in series. OPEN means that R is infin-
ite so that the standard behaves like a capacitor (no induc-
tance, the polynomial coefficients specify C). SHORt means
that R is zero so that the standard behaves like an inductance
(no capacitance, the polynomial coefficients specify L). MATCh
means that the standard behaves like a match with the speci-
fied <Resistance>
Character data
<Resistance>
Optional resistance for the MATCh standard
Numeric value
, <Port1>, <Port2>
Optional port restriction: one port number for one port stand-
ards, two port numbers for two port standards
Integer value
SCPI Command Reference