Command Reference
R&S
®
ZNC
468
User Manual 1173.9557.02 ─ 13
CONTrol:HANDler:PASSfail:SCOPe
<Scope>
Specifies the "sweep end" condition that will cause the /PASS FAIL line (pin 33) to report
the status of the global limit check.
Note:
This setting is not valid if the pass/fail mode is NOWait (
NOWait
).
Parameters:
<Scope>
GLOBal | CHANnel
CHANnel
–
when all the sweeps for each channel are complete
GLOBal
–
when all sweeps in all channels are complete
Note:
*RST
or "Preset" do not change the configuration of the Uni-
versal Interface. Use
default values.
*RST:
n/a (default: GLOBal)
Example:
See
CONTrol:HANDler:PASSfail:STATus?
Returns the global pass/fail status of the last measurement.
Return values:
<Status>
PASS | FAIL | NONE
PASS
–
all measurements that are not in single sweep mode (on
hold) have been swept, and all limit checks have been passed.
FAIL
–
all measurements that are not in single sweep mode (on
hold) have been swept, at least one limit check failed according to
the specified pass/fail policy (
NONE
–
no pass/fail status available, e.g. because the measure-
ment is in progress or because no limit check has been defined.
*RST:
n/a
Example:
Preparations: Configure and enable a limit check. Start a mea-
surement and wait until the sweep is complete.
CONTrol:HANDler:PASSfail:STATus?
Query the result of the global limit check.
Usage:
Query only
CONTrol:HANDler:RESet
Restores the default states of the
CONTrol:HANDler...
commands including the data
port values.
Example:
See
Usage:
Event
SCPI Command Reference