Command Reference
R&S
®
ZNC
631
User Manual 1173.9557.02 ─ 13
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the param-
eter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parameter
is valid for all sources.
Parameters:
<NoReadings>
Number of power meter readings. If the value exceeds the maxi-
mum number of calibration sweeps
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
AVERage[:COUNt]
) the latter is adjusted.
Range:
1 to 100
*RST:
1
Example:
See
.
SOURce<Ch>:POWer:CORRection:DATA
<Source>, <CorrData>...
Reads or writes scalar source power correction data sets. A power correction data set
contains n real values where:
●
Each value corresponds to the ratio of the actual power at the reference plane (value
provided by the used source) to the uncalibrated power in dB.
●
The number n is equal to the number of sweep points.
Increasing (decreasing) the values in the correction data sets increases (decreases) the
power at the reference plane. Writing correction data (the setting command) fails if the
number of transferred values is not equal to the number of sweep points.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Source>
Identifier for the source of the calibrated wave:
'
A<n>
' or '
PORT<n>
' denote correction data for the analyzer port
no.
<n>
.
<CorrData>
<block_data>
Power correction values in block data format or as a comma-sep-
arated list of real numbers. The query returns a list of numbers.
Range:
All numbers are interpreted as dB-values.
SCPI Command Reference