The bandfilter search results are displayed in the movable bandfilter info field.
Remote command:
BPASs
BFILter
Bandpass Ref to Mkr
Activates the search for a bandpass region on the active trace and activates bandfilter
"Tracking", starting at the position of the active marker. A bandpass region is the closest
peak in the search range with a minimum excursion as specified by the "Bandwidth"
parameter.
In contrast to a "Bandpass Ref to Max", the "Bandpass Ref to Mkr" does not change the
position of the active markers. The bandfilter search results are displayed in the movable
bandfilter info field.
Remote command:
BPRMarker
BFILter
Bandstop Ref to Max
Activates the search for a bandstop region on the active trace and activates bandfilter
"Tracking". A bandstop region is the lowest peak (local minimum) in the search range,
provided that its level is at least <Bandwidth> below the maximum (passband value).
If "Bandstop Ref to Max" is selected the analyzer uses (or creates) the four markers
"M1" to "M4" to locate the
bandstop region
.
●
"M1" indicates the minimum of the peak ("Min").
●
"M2" indicates the point on the left edge of the peak where the trace value is equal
to the maximum in the search range (passband value) minus the bandwidth factor
("Lower Edge").
●
"M3" indicates the point on the right edge of the peak where the trace value is equal
to the maximum in the search range (passband value) minus "x dB Bandwidth" (Upper
Edge).
●
"M4" indicates the center of the peak, calculated as the geometric mean value of the
"Lower Edge" (LBE) and "Upper Edge" (UBE) positions:
Trace Settings