GUI Reference
R&S
®
ZNC
223
User Manual 1173.9557.02 ─ 13
Tip: Generator power
The power range for a power sweep replaces the fixed internal source power which is
defined via "POWER BW AVG > Power". The power is the output power at all test ports
which supply the stimulus signals for the active channel.
Remote command:
POWer
CW Mode
"CW Mode" sweeps, like "Time" sweeps, are performed at constant frequency and stim-
ulus power. The measurement is triggered according to the current trigger settings; each
trigger event triggers the first partial measurement of a measurement point. The time
interval between two consecutive measurements depends on the trigger settings and the
sweep parameters (especially the number of points). Any trigger mode is allowed.
The frequency ("CW Frequency") and internal source power ("Power") are fixed. Both
can be defined in the "CHANNEL > SWEEP > Sweep Params" tab.
A "CW Mode" sweep corresponds to the analysis of a signal over the time with a time
scale and resolution that is determined by the trigger events. In a Cartesian diagram the
measurement result is displayed as a trace over a linear time scale (like e.g. in an oscil-
loscope). The diagram is similar to the "Time" diagram. The following example shows a
"CW Mode" sweep with a DUT that does not markedly change its transmission charac-
teristics over the time.
Tip: Sweep time
Channel Settings