Command Reference
R&S
®
ZNC
618
User Manual 1173.9557.02 ─ 13
Example:
SWE:TIME 1
Set a total sweep time of 1 s.
SWE:TIME:AUTO?
A query returns the value
1
.
Manual operation:
See
[SENSe<Ch>:]SWEep:TYPE
<Format>
Selects the sweep type, i.e. the sweep variable (frequency/power/time) and the position
of the sweep points across the sweep range.
Suffix:
<Ch>
.
Channel number.
Parameters:
<Format>
LINear | LOGarithmic | POWer | CW | POINt | SEGMent
LINear - linear frequency sweep at constant source power (
(
[SENSe<Ch>:]FREQuency:...
) is swept in equidistant steps
over the frequency range. In a Cartesian diagram, the x-axis is a
linear frequency axis.
LOGarithmic - logarithmic frequency sweep. The frequency is
swept in equidistant steps on a logarithmic scale. In a Cartesian
diagram, the x-axis is a logarithmic frequency axis.
SEGMent - segmented frequency sweep. The sweep range is
composed of several continuous frequency ranges or single fre-
quency points defined by means of the commands in the
[SENSe<Ch>:]SEGMent<Seg>...
subsystem.
POWer - power sweep. The measurement is performed at con-
stant frequency (
but with variable generator power that is swept in linear, equidis-
tant steps over a continuous range
(
SOURce<Ch>:POWer<PhyPt>:STARt
). In a Cartesian diagram, the x-axis is a
dB-linear power axis.
CW - time sweep. The measurement is performed at constant fre-
quency (
) and repeated over a specified
period of time at constant time intervals.
POINt - CW mode sweep, time sweep triggered according to the
current trigger settings.
*RST:
LINear
SCPI Command Reference