Command Reference
R&S
®
ZNC
561
User Manual 1173.9557.02 ─ 13
<Characterization>
Name and (possibly) directory of the characterization (cal kit file)
to be used for the automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file stored
in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path denotes a spe-
cific cal kit file stored in the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path denotes a specific
cal kit file stored in an arbitrary directory.
<TestPort1>
Test port numbers. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified. For
a one path two port calibration (OPTPort), the first port no. denotes
the node port (fully calibrated port).
<TestPort2>
Second test port number of the analyzer
Example:
CORR:COLL:AUTO:TYPE FNPort, '', 1, 2, 4
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and arbitrary
test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:COLLect:AVERage
<Average>
Activates automatic averaging, which means that the VNA may perform multiple calibra-
tion sweeps and apply averaging to reduce trace noise. In contrast to manual averaging
(see
) the number of calibration sweeps is calcula-
ted automatically.
Parameters:
<Average>
AUTO | MANual
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:ALL
<Boolean>
Enables calibration of all channels in the active recall set. The command is valid for the
following calibration methods:
●
Manual system error correction
●
Automatic system error correction (cal unit)
A scalar power calibration is not affected.
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
SCPI Command Reference