GUI Reference
R&S
®
ZNC
310
User Manual 1173.9557.02 ─ 13
Preset Configuration
Specifies whether "System >Preset" will perform a factory preset or restore the settings
stored in a user preset file. A user preset file is an arbitrary recall set (
.znx
) file, to be
stored using "FILE > Setup > Save...". If the current user preset file is not found (e.g.
because it was deleted or moved), "System >Preset" initiates a factory preset.
Remote command:
4.6.2.2
Calibration
Provides general system error correction (calibration) settings.
Auto Power Reduction for Cal Unit
Sets the source power at all test ports to –10 dBm while an automatic calibration using
the calibration unit R&S ZV-Z5x is active. Applying this source power to the ports of the
calibration unit ensures best accuracy of the automatic calibration. The source powers
are reset to their original values after the calibration is completed. The automatic power
reduction can be deactivated in case that the test setup introduces a large attenuation.
Remote command:
Auto Averaging
Activates automatic averaging, which means that the VNA may perform multiple calibra-
tion sweeps and apply averaging to reduce trace noise. In contrast to regular averaging
(see
chapter 4.4.1.3, "Power Bw Avg > Average"
, on page 216) the number of calibration
sweeps is calculated automatically.
Remote command:
System Settings