Command Reference
R&S
®
ZNC
654
User Manual 1173.9557.02 ─ 13
Parameters:
<CalUnitName>
String parameter containing the name of a cal unit characteriza-
tion.
'Factory'
denotes the factory characterization; an empty
string
' '
denotes the last referenced characterization.
Example:
See
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction[:STATe]
<Boolean>
Enables or disables automatic power reduction at all test ports while an automatic cali-
bration using the calibration unit R&S ZV-Zxx is active.
Parameters:
<Boolean>
Power reduction enabled or disabled.
*RST:
ON
Example:
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction OFF
Disable automatic power reduction.
Manual operation:
See
"Auto Power Reduction for Cal Unit"
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
<CalUnitName>, <Type>,
<SParameter>[, <FirstPort>, <SecondPort>]
Reads the calibration data for a particular standard from a cal unit characterization.
Parameters:
<CalUnitName>
String parameter containing the name of a cal unit characteriza-
tion.
'Factory'
denotes the factory characterization; an empty
string
' '
denotes the last referenced characterization.
<Type>
THRough | OPEN | SHORt | MATCh | MMTHrough | MFTHrough |
FFTHrough | MOPen | FOPen | MSHort | FSHort | MMTCh |
FMTCh
Standard types; for a description refer to table
. The standard types of a particular characteri-
zation can be queried via
The factory characterization usually does not contain data for a
through standard; therefore a query of the type
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
'Factory', THR, S11, 1, 2
results in an error message.
<SParameter>
S11 | S12 | S21 | S22
S-parameter of the standard, use
S11
for one-port standards, one
of the four for 2-port standards.
SCPI Command Reference