Command Reference
R&S
®
ZNC
535
User Manual 1173.9557.02 ─ 13
Usage:
Event
Manual operation:
See
[SENSe<Ch>:]AVERage:COUNt
<AverageFactor>
Defines the number of consecutive sweeps to be combined for the sweep average
("Factor").
Suffix:
<Ch>
.
Channel number
Parameters:
<AverageFactor>
Sweep average factor
Range:
1 to 1000
*RST:
10
Example:
See
Manual operation:
See
[SENSe<Ch>:]AVERage:MODE
<Mode>
Suffix:
<Ch>
.
Parameters:
<Mode>
AUTO | FLATten | REDuce
AUTO
Automatic selection of the averaging mode, depending on the
trace format.
FLATten
Averaging of the (linear) magnitude and phase values, provides
the most effective noise suppression for the "dB Mag", "Phase",
"Unwr. Phase", and "Lin Mag" formats.
REDuce
Averaging of the real and imaginary parts of each measurement
result, provides the most effective noise suppression for the
"Real" and "Imag" formats and for polar diagrams.
Manual operation:
See
[SENSe<Ch>:]AVERage[:STATe]
<Boolean>
Enable or disable the sweep average.
Suffix:
<Ch>
.
Channel number
SCPI Command Reference