218
User Manual 1173.9557.02 ─ 13
Background information
Refer to the following sections:
●
chapter 3.1.4, "Sweep Control"
●
chapter 3.1.4.2, "Stimulus and Sweep Types"
4.4.2.1
Sweep > Sweep Params
Controls the scope and timing of the measurement in the active channel.
Segmented sweeps
In "Segmented Frequency" sweeps, the sweep parameters can be set independently for
each segment. see
chapter 4.4.2.3, "Define Segments (Dialog)"
System error correction
In general, the system error correction is no longer valid after a change of the sweep
parameters. The status of the calibration is shown in the trace list. If the number of points
is changed, the analyzer interpolates the correction data. The calibration label "Cal Int"
is displayed. See also
chapter 3.5.4, "Calibration Labels"
Access:
CHANNEL > SWEEP key or Alt + Shift + M
Number of Points
Sets the total number of measurement points per sweep. The minimum number of points
is 1 (measurement at a single frequency/power/time value). The maximum depends on
the analyzer type.
Together with the sweep range defined with the "Stimulus" settings, this parameter
defines the grid of sweep points. The sweep points are equidistantly distributed over the
entire sweep range: The step width between two consecutive sweep points is constant
on a linear scale (sweep types "Lin. Frequency", "Time" and "CW Mode") or on a loga-
rithmic scale (sweep types "Log. Frequency" and "Power").
Channel Settings