Command Reference
R&S
®
ZNC
622
User Manual 1173.9557.02 ─ 13
<SweepType>
CW | FIXed | SWEep
SWEep - the full formula is applied. For frequency sweeps, the
command defines a sweep range.
CW | FIXed - the reduced formula with f
b
= 0 is applied; the com-
mand defines a fixed frequency.
*RST:
SWEep
Example:
*RST; FREQ:STAR 1E+9; STOP 1.1E+9
Reset the analyzer (activating a frequency sweep) and set the
sweep range between 1 GHz and 1.1 GHz.
SOUR:FREQ2:CONV:ARB:IFR 2, 1, 1E+9, SWE
Convert the source frequency at test port no. 2 to the range
between 3 GHz and 3.2 GHz.
SENSe:FREQuency:CONVersion:GAIN:LMCorrection ON
Enable the load match correction for all conversion gains.
SENSe:FREQuency:CONVersion?
Query the measurement mode. The response is
ARB
, because the
port frequency definition also activates the arbitrary mode.
SENSe:FREQuency:CONVersion FUND
De-activate the arbitrary mode, cancel the port frequency defini-
tion.
SOURce<Ch>:FREQuency<PhyPt>[:CW]
<FixedFreq>
SOURce<Ch>:FREQuency<PhyPt>:FIXed
<FixedFreq>
Defines the fixed (Continuous Wave, CW) frequency for all sweep types operating at fixed
frequency (power sweep, time sweep, CW mode sweep). The two command forms
SOURce<Ch>:FREQuency<PhyPt>:CW
and
SOURce<Ch>:FREQuency<PhyPt>:FIXed
are equivalent.
The frequency range is listed in
.
Note:
SOURce<Ch>:FREQuency<PhyPt>[:CW]|:FIXed
is equivalent to
[SENSe<Ch>:]FREQuency[:CW]|:FIXed
. Source and receiver frequency are always
equal; the four commands overwrite each other.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected frequency applies to all source ports in the active
channel.
Parameters:
<FixedFreq>
Fixed stimulus and analyzer frequency.
Increment: 0.1 kHz
*RST:
1 GHz
Default unit: Hz
SCPI Command Reference