GUI Reference
R&S
®
ZNC
262
User Manual 1173.9557.02 ─ 13
The properties of the selected characterization are shown below the list. "Delete" deletes
the selected characterization file; "Start Characterization" opens the characterization
wizard to create a new characterization.
Remote command:
4.4.3.11
Characterization Wizard
The "Characterization" wizard guides you through the automatic characterization of a
calibration unit.
The guided characterization consists of the following steps:
1.
Characterization:
Select the characterized ports and cal unit standards and initiate
the characterization sweeps.
2.
Save Characterization Data:
Save the characterization data to an internal file on
the calibration unit.
Step 1: Characterization
Selects the calibration type as well as the characterized cal unit ports and initiates the
necessary characterization sweeps.
Characterization procedure
To acquire accurate characterization data, the test setup must be properly calibrated
before you start the characterization wizard. Use the calibration type that you wish to
perform with your new cal unit characterization; see
chapter 3.5.5.3, "Characterization of
Access:
CHANNEL > CAL > Cal Devices > Characterize Cal Unit... > Start Characteri-
zation...
Channel Settings