Command Reference
R&S
®
ZNC
408
User Manual 1173.9557.02 ─ 13
<Mk>
Marker number in the range 1 to 10. For a bandfilter search
(
BFILter
) this numeric suffix is ignored and may be set to any
value because the bandfilter search functions always use markers
M1 to M4.
Setting parameters:
<SearchMode>
MAXimum | MINimum | RPEak | LPEak | NPEAK | TARGet |
LTARget | RTARget | BFILter
See list of parameters below.
Example:
Suppose that the active recall set contains an active trace no. 1.
CALC:MARK ON
Create marker M1 and assign it to trace no. 1.
CALC:MARK:FUNC:EXEC MAX; RES?
Move the created marker to the absolute maximum of the trace
and query the stimulus and response value of the search result.
Usage:
Setting only
Manual operation:
See
The analyzer provides the following search modes:
Mode
Find...
MAXimum
Absolute maximum in the search range (see
MINimum
Absolute maximum in the search range
RPEak
Next valid peak to the right of the current marker position
LPEak
Next valid peak to the left
NPEak
Next highest or lowest value among the valid peaks (next peak)
TARGet
RTARget
Next target value to the right of the current marker position
LTARget
Next target value to the left
BFILter
Bandfilter search. The results are queried using
CALCulate<Chn>:MARKer<Mk>:FUNCtion:RESult?
Returns the result (stimulus and response value) of a search started by means of
. The search must be executed
before the command is enabled.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10. This numeric suffix is ignored
and may be set to any value.
Example:
See
SCPI Command Reference