background image

587M

6400

Automatic Isoperibol Calorimeter

Operating Instruction Manual

For models produced after October 2010

Содержание 6400

Страница 1: ...587M 6400 Automatic Isoperibol Calorimeter Operating Instruction Manual For models produced after October 2010...

Страница 2: ......

Страница 3: ...ocess 22 Closing the bomb 22 Cool Rinse 26 Drain Fill Cycle 23 Pre Period 24 Bomb Firing 25 Post Period 25 Chapter 4 28 Menu Descriptions 28 Main Menu 28 Calorimeter Operation Menu 28 Temperature vs T...

Страница 4: ...Server Feature Optional 59 Bar Code Port 66 Chapter 9 68 Memory Management 68 Clearing Memory 68 Removable SD Memory 68 Chapter 10 70 Maintenance 70 Inspection of Critical Sealing Surfaces 70 Bomb Re...

Страница 5: ...gure 14 10 6400 Water Tank and Jacket Cooling Solenoid 91 Tables Table 6 1 Calorimeter Control Limit Values in J g When Benzoic Acid is Used as a Test Sample 49 Table 6 2 Calorimeter Control Limit Val...

Страница 6: ...his book No Description 201M Limited Warranty 207M Analytical Methods for Oxygen Bombs 230M Safety in the Operation of Laboratory and Pressure Vessels 483M Introduction to Bomb Calorimetry Note The un...

Страница 7: ...t specified by Parr Instrument Company the protection pro vided by the equipment may be impaired Explanation of Symbols I On Position O Off Position Alternating Current This CAUTION symbol may be pres...

Страница 8: ...ditions Operating 15 C to 30 C maximum relative humid ity of 80 non condensing Installation Category II over voltage in accordance with IEC 664 Pollution degree 2 in accordance with IEC 664 Altitude L...

Страница 9: ...w w w p a r r i n s t c o m 7 Notes 6400...

Страница 10: ...rain SwagelokTube Fittings When SwagelokTube Fittings are used the instruc tions for installation are 1 Simply insert the tubing into the SwagelokTube Fitting Make sure that the tubing rests firmly on...

Страница 11: ...r are required to fill the external pressurized rinse tank Approximately 2L of distilled de ionized or tap water with a total hardness of 85 ppm or less are required for filling the internal cooling r...

Страница 12: ...1 Installation P a r r I n s t r u m e n t C o m p a n y 10 Figure 1 2 6400 Calorimeter Back Panel...

Страница 13: ...l initially accept about 2 liters Fill the external rinse tank with about 16 liters of distilled water through the large opening at the top Figure 1 3 6400 External Plumbing 06 05 07 HJA of the tank T...

Страница 14: ...en filling system Keep all threads fit tings and gaskets clean and in good condition Note To release the pressure inside the rinse tank turn off the gas supply and open the gas relief valve lever Gas...

Страница 15: ...the Network Data Device menu in the Communications Controls menu Balance Interface devices are polled from device 1 to 15 for sample and or spike weights when the weight entry mode is set to Network...

Страница 16: ...1 Installation P a r r I n s t r u m e n t C o m p a n y 14 Figure 1 4 6400 Calorimeter Peripherals Figure 1 5 Multiple Alternate Configurations...

Страница 17: ...w w w p a r r i n s t c o m 15 Notes 6400 1...

Страница 18: ...head in the calorimeter 6 Close the calorimeter cover making sure that the latch is engaged 7 Select determination or standardization as ap propriate on the Calorimeter Operation page by toggling the...

Страница 19: ...w w w p a r r i n s t c o m 17 Notes 6400 2...

Страница 20: ...nd the new value will be displayed in the lower right corner of the key 4 Data Displays Most of these keys display values that have been calculated by the calorimeter and are informational only Certai...

Страница 21: ...etting go to the Pro gram Info and Control Page User Factory Settings Re load User Default Settings andYES Sample Preparation Sample Size To stay within safe limits the bomb should never be charged wi...

Страница 22: ...us and fluffy materials generally require one of three modes for controlling the burn rate Fibrous materials do not pelletize readily and generally require either moisture content or a combustion aid...

Страница 23: ...lowing procedure when filling and han dling any of these tape sealed sample holders 1 Weigh the empty cup or capsule then cover the top with tape trim with a knife and press the trimmed edge firmly ag...

Страница 24: ...or 43A3KQ Fused Quartz is required Test Process Loading the sample Prepare and weigh the sample to 0 0001g Gently tap capsules that contain powdered samples to compact the material Pellets are easier...

Страница 25: ...ty 2 The water fill solenoid opens and water is pumped from the internal tank into and through the bucket that surrounds the bomb Overflow from the bucket is returned to the closed water tank Because...

Страница 26: ...e system Water within this bucket is circulated by the stirrer Water continues to circulate from the closed water system through the jacket surrounding the bucket 2 The oxygen filling valve closes and...

Страница 27: ...period A warning of short beeps is sounded indicating the bomb is about to be fired Post Period A minimal temperature rise will confirm that the sample has ignited After this verification the post per...

Страница 28: ...ed and the residual pressure is released through the bomb ex haust line 3 Once the excess oxygen is vented the bomb rinse water from the rinse water tank is admitted through the bomb rinse solenoid va...

Страница 29: ...the bomb rinse sequence the drain sequence begins Figure 3 6 Drain Flow Diagram The water in the bucket is drained out of the bucket and routed to the drain connection Once the bucket is drained the...

Страница 30: ...st or pretest in progress Calorimeter Operation Menu The Calorimeter will normally be operated from the Calorimeter Operation Menu although tests can always be started from any menu screen Operating M...

Страница 31: ...minimum bucket value Jacket Plot Symbol Toggles between same as Bucket Plot Symbol above Jacket Min Value Press this key to access its numeric dialog box to set a minimum jacket value Time Window Sets...

Страница 32: ...on Spiking Controls Use Spiking When set to ON the calorimeter will prompt for the weight of the spike added and will compensate for the heat of combus tion in the calculations Heat of Combustion of S...

Страница 33: ...begins The factory default value is 2s ClearTime This time value is used to establish a post rinse oxygen filling time for the bomb This step is used to clear the lines and valves of any residual rin...

Страница 34: ...ens a window with a slide adjustment to set the volume of the key cliks and alarms of the calorimeter Default is 85 Software and Hardware Info This screen displays important information such as the ma...

Страница 35: ...y once the user has config ured the instrument to their operating require ments Compare Settings With Factory Defaults This button will bring up a screen that will show the differences in the current...

Страница 36: ...ntrol chart is a graphical tool which can assist the user in deter mining whether or not their process is in control Many standard methods will dictate that a reference sample be measured periodically...

Страница 37: ...the Calibration Data and Controls menu If the Protect EE value is set to ON this key is not functional Manual EE Entry This key allows the user to manually enter an EE or calibration factor for a give...

Страница 38: ...dialog box on which the value can be set An appropriate Fixed HNO3 value is 8 calories when one gram benzoic acid pellets are used to calibrate the instrument Fixed Sulfur Correction Press this key o...

Страница 39: ...used for the sulfur correction are multiplied by this value to get the product into units of milliequivalents The default number 0 6238 requires that the sulfur value be entered in weight percent Fuse...

Страница 40: ...nd set the value Units are weight Fixed Moisture as Received Press the LEFT side to toggle ON or OFF whether to use the entered moisture correction Press the RIGHT side to access the Fixed Moisture as...

Страница 41: ...fix for all sample IDs if the Auto matic Sample ID is set to ON Press this key to access a sub menu for entering an alphanu meric prefix Next Auto Sample ID Number Establishes the initial sample numbe...

Страница 42: ...idth of the printer to either 40 or 80 columns Select 40 when the 1758 Printer is used Automatic Reporting Toggles the automatic report ing ON OFF When ON preliminary reports will be generated at the...

Страница 43: ...d rate and balance type are provided to match any devices that might be connected to these ports Balance Port Settings Number of Data Bits Standard options for data bits Toggles between 7 and 8 Parity...

Страница 44: ...Accesses a submenu to set up a Bar Code Scanner for use with the calorimeter Network Data Devices Accesses a submenu to input the IP addresses of networked devices such as balances and proximate analy...

Страница 45: ...in blue Desel All This key deselects all files previously selected Rename This key allows the user to rename the blue highlighted filename Delete This key deletes all selected files ConvertType This k...

Страница 46: ...ll records are date and time stamped Helpful items to log are D0 Corrected calorimeter drift rate Tsum Accumulated temperature rise T1 Extrapolated temperature rise C0 Temperature conversion counter D...

Страница 47: ...is key to display a screen with current operating system information statistics such as Processes and their associated PIDs Memory Mass Storage Network Press the Print key to print this information Vi...

Страница 48: ...ly pre weighed samples reports all stored reports or preliminary and final reports Prompt For FinalValues When turned on the controller will prompt the operator to enter any missing corrections for fu...

Страница 49: ...w w w p a r r i n s t c o m 47 Notes 6400 5...

Страница 50: ...ellets Part No 3415 is furnished with each calorimeter for standardizing purposes The Parr benzoic acid has been calibrated against NIST benzoic acid Addition al benzoic acid pellets can be obtained f...

Страница 51: ...ter a value appropriate for the test being made Table 6 1 Calorimeter Control Limit Values in J g When Benzoic Acid is Used as aTest Sample Accepted heat of combustion taken as 26454 J g Instrument pr...

Страница 52: ...WITHIN THE GROUP UCL FORTHE RSD WITHIN THE GROUP MAXIMUM PERMISSIBLE DEVIATION OFTHE GROUP MEAN FROMTHE ACCEPTED VALUE OR GRAND MEAN 1 19 0 2 23 3 0 261 13 4 3 27 5 0 228 10 9 4 29 7 0 209 9 5 5 31 1...

Страница 53: ...THE GROUP UCL FORTHE RSD WITHIN THE GROUP MAXIMUM PERMISSIBLE DEVIATION OFTHE GROUP MEAN FROMTHE ACCEPTED VALUE OR GRAND MEAN 1 34 1 2 41 9 0 261 24 1 3 49 6 0 228 19 7 4 53 4 0 209 17 1 5 55 9 0 196...

Страница 54: ...in the formula or abbreviated as EE is determined by standardizing the calorimeter as described in Chap ter 6 Standardization It is an expression of the amount of energy required to raise the tempera...

Страница 55: ...before it ignites Using the fine thread mentioned above the fuse correction for the calorimeter would be the 10 calories from electrical heating plus 12 calories from the burning thread for a total of...

Страница 56: ...mple mass is the mass of sample burned in the bomb grams Sulfur multiplier is multiplied by the product of the user entered sulfur value and the sample mass to arrive at the number of milliequivalents...

Страница 57: ...plied to samples containing up to 2 nitrogen without introducing a significant error in the resulting heat of combustion value Users may find it convenient to enter a fixed value for the acid correcti...

Страница 58: ...lculations Both the ISO 1928 and BSI 1016 Part 5 methods for testing the calorific value of coal and coke deal with acid and sulfur corrections in a manner which is somewhat different than ASTM proced...

Страница 59: ...bove give the calorific value of the sample with moisture as it existed when the sample was weighed For example if an air dried coal sample was tested the results will be in terms of heat units per we...

Страница 60: ...nected to LAN side of the router The PC in turn is also connected to the LAN side of the router using a similar CAT 5 cable A D Link 614 router is recommended for this purpose For this router operated...

Страница 61: ...eature can be used to facilitate data file transfer from a calorimeter or proximate interface to a PC running the Windows operating system This method of file transfer for some users may be less cumbe...

Страница 62: ...n y 60 The calorimeter offers a web server service Test reports can be viewed with a web browser using a URL of the following form http 10 1 5 10 Where 10 1 5 10 is the IP address of the calorimeter...

Страница 63: ...ComputerCommunications 6400 8 w w w p a r r i n s t c o m 61 Clicking on the Config button will display the screen below Changes made on this screen will change the settings in the calorimeter...

Страница 64: ...r I n s t r u m e n t C o m p a n y 62 Clicking on the Run Data button displays a list of reports currently in the instrument memory Clicking on a test under the select sample ID box will display the...

Страница 65: ...ComputerCommunications 6400 8 w w w p a r r i n s t c o m 63 Clicking on the System Info button will display the screen below...

Страница 66: ...screen displayed by the calorimeter If the backlight is not on this screen will display a blank blue square Note This is a picture only The calorimeter cannot be remotely operated from this screen Re...

Страница 67: ...8 w w w p a r r i n s t c o m 65 Clicking on the Documentation button will display the screen below Clicking on any of the links will open the corresponding web page Note Connection to the internet i...

Страница 68: ...eature Key and enter the feature key purchased from Parr Instrument Company into the instrument by using the touchpad Pressing the key labeled ABC allows the user to switch from upper case letters to...

Страница 69: ...Hydrogen hydrogen value net calc option HydrogenFinal hydrogen value is final net calc option MAD moisture as determined value dry calc option MAD Final moisture as determined is final JacketTemp jack...

Страница 70: ...troller will then ask the user to confirm that this file is to be deleted A series of files can be deleted by selecting the first file in the series and then the last file in the series using the EXTE...

Страница 71: ...w w w p a r r i n s t c o m 69 Notes 6400 9...

Страница 72: ...bombs with significant imper fection of this nature be serviced at Parr CAUTION Do not pry elastomer seals O rings and quad rings from seal grooves with metallic tools The use of dental picks and othe...

Страница 73: ...ter Clean the grill on the heat sink at the rear of the calorimeter Periodic cleaning may be performed on the exterior surfaces of the instrument with a damp cloth All power should be disconnected whe...

Страница 74: ...s oriented toward the left side of the instrument The insert is keyed to the cylinder and can not be fully inserted unless it is properly oriented 5000Test Maintenance To deal with the realities of to...

Страница 75: ...Date Date Date Date 1 Replace 8400DD2 Heating Wire 2 Clean 986DD Electrode Contact Pins 500Test Maintenance Replace the following Date Date Date Date Head 1374HCJV 2 O rings 394HC O ring 694DD O ring...

Страница 76: ...testing purposes Be aware that all user and instrument protection is disconnected while on this screen This is very important and you should take proper precaution 1 Make sure the 668DD check valve i...

Страница 77: ...cated For reference purposes the normal upward thrust generated by the 966DD2 piston is 50 pounds The downward thrust is 20 pounds Far less than 20 pounds are required to move the piston in either dir...

Страница 78: ...he bomb within the time allowed Possible causes for this error are listed below A bomb leak Poor bucket stirring Lid not tight A Postperiod Timeout has Occurred The calorimeter has failed to establish...

Страница 79: ...detected The heater will now be shutdown This error means that the calorimeter is trying to heat the water in the unit for an extended period of time When the heater and pump are initially turned on t...

Страница 80: ...e it is helpful if the person is close to the instrument in order to implement any changes recommended by theTechnical Service Department Return for Repair To return the instrument for repair please c...

Страница 81: ...w w w p a r r i n s t c o m 79 Notes 6400 12...

Страница 82: ...e Heater Assembly 230V 139E23 Fuse Fast Act 15 AMP 250V Caution For continued protection against possible hazard replace fuses with same type and rating of fuse A1250DD2 Controller Assembly Item Descr...

Страница 83: ...ing 0 125 ID SN1332HX 6 32 Hex Nut 18 8 SS A1400DD Plunger Knob Assembly 1218DD Cover Air 1229DD3 Plate 1237DD2 Friction Hinge 1396DD Latch Block 1324DD2 Post Latch Locking A1230DD2 Cover Tubing Assem...

Страница 84: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 82 Chapter 14 Drawings Figure 14 1 1138 Parts Diagram...

Страница 85: ...DDJB O ring 1 5 8 ID NBR 21 1452DD Head 1452DDCL Head for Chlorine Service 22 655DD 2 Electrode Spacer 23 1095DD Electrode 24 PA1332RD04 2 6 32 x 1 4 RHMS 25 656DD Reducer Bushing 26 653DD Electrode N...

Страница 86: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 84 Figure 14 2 6400 Cutaway Right...

Страница 87: ...Drawings 6400 14 w w w p a r r i n s t c o m 85 Figure 14 3 6400 Cutaway Left...

Страница 88: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 86 Figure 14 4 6400 Cover Open 1324DD2 Locking Post...

Страница 89: ...F CPU A1822E POWER CABLE ASSY A2166E IO TO CPU USB CABLE ASSY RED WIRE ON CABLE CONNECTS TO PIN 1 A2154E CPU BOARD A2165E REF A BY BY SCALE DWG NO DATE DRAWN DESCRIPTION REVISIONS FOR DATE CONTROLLER...

Страница 90: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 88 Figure 14 6 A1251DD Oxygen Solenoid Assembly Figure 14 7 A1447DD Water Solenoid Assembly...

Страница 91: ...Drawings 6400 14 w w w p a r r i n s t c o m 89 Figure 14 8 A1456DD Rinse Valve Assembly...

Страница 92: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 90 Figure 14 9 6400 Internal Plumbing Diagram...

Страница 93: ...Drawings 6400 14 w w w p a r r i n s t c o m 91 Figure 14 10 6400 Water Tank and Jacket Cooling Solenoid...

Страница 94: ...I n s t r u m e n t C o m p a n y 92 Figure 14 11 A1455DD Propeller Assembly Apply thread sealant Locktite or equivalent to set screw in 1242DD2 pulley before installing Figure 14 12 A1448DD Temperatu...

Страница 95: ...Drawings 6400 14 w w w p a r r i n s t c o m 93 Figure 14 13 A1268DD Stirrer Motor and Mount...

Страница 96: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 94 Figure 14 14 6400 Bucket Assembly...

Страница 97: ...Drawings 6400 14 w w w p a r r i n s t c o m 95 Figure 14 15 6400 Air Can Assembly Cutaway Left...

Страница 98: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 96 Figure 14 16 6400 Air Can Assembly Cutaway Front...

Страница 99: ...Drawings 6400 14 w w w p a r r i n s t c o m 97 Figure 14 17 A1450DD Bomb Head Assembly View 1...

Страница 100: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 98 Figure 14 18 A1450DD Bomb Head Assembly View 2...

Страница 101: ...051DD EXHAUST DIFFUSER 308VBAD MALE CONNECTOR 1 4T 1 4NPTM HX0025TB062 1 4 TUBING X 4 FT LG 411VBAD REDUCER 1 4T 3 8Tx 221VBAD 3 8T NUT 222VBAD 3 8T FERRULE SET SEE NOTE 1 NOTES 1 CONNECTS TO VESSEL E...

Страница 102: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 100 Figure 14 20 Wiring Diagram...

Страница 103: ...Drawings 6400 14 w w w p a r r i n s t c o m 101 Figure 14 21 Wiring Diagram...

Страница 104: ...14 Drawings P a r r I n s t r u m e n t C o m p a n y 102 Figure 14 22 Fuse Diagram...

Страница 105: ...w w w p a r r i n s t c o m 103 Notes 6400 14...

Страница 106: ...nseTank Empty ON RinseTank Capacity 150 Rinses Left 150 Reset RinseTank Counter RinseTime 25 Rinse FlushTime 20 ClearTime 100 of Rinse Cycles 3 Program Information and Controls Date XX XX XXXX Time XX...

Страница 107: ...ls Hydrogen Entry Mode Touch Screen Oxygen Entry Mode Touch Screen Nitrogen Entry Mode Touch Screen Moisture Data Entry Controls MAD Entry Mode Touch Screen MAR Entry Mode Touch Screen Auto Sample ID...

Страница 108: ...Line Setting Value Thermochemical Corrections Acid Correction STD Entered HNO3 13 Fixed Sulfur STD Off 7 Acid Correction DET Entered HNO3 13 Fixed Sulfur DET Off 7 Calculations Factors Acid Multiplie...

Страница 109: ......

Страница 110: ...587M R03 02 11 14...

Отзывы: