Command Reference
R&S
®
ZNC
406
User Manual 1173.9557.02 ─ 13
Example:
See
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:CENTer
Sets the center of the sweep range equal to the stimulus value of the marker
<Mk>
on
trace no.
<Chn>
.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Example:
*RST; :CALC:MARK ON
Create marker 1 in the center of the current sweep range and
assign it to trace no. 1.
CALC:MARK:FUNC:CENT
Leave the sweep range unchanged.
Usage:
Event
Manual operation:
See
CALCulate<Chn>:MARKer<Mk>:FUNCtion:DOMain:USER
<NumSearchRange>
Assigns a search range no.
<NumSearchRange>
to marker no
<Mk>
and selects the
search range, e.g. in order to display range limit lines or define the start and stop values.
Suffix:
<Chn>
.
Channel number used to identify the active trace.
<Mk>
Marker number in the range 1 to 10.
Parameters:
<NumSearchRange> Number of the search range.
Range:
0 to 10. 0 denotes the fixed full span search range
(equal to the sweep range), 1 to 10 denote user-
definable search ranges; see example.
*RST:
0 (reserved for full span search range)
Example:
CALC1:MARK1:FUNC:DOM:USER 2
Select the search range no. 2, assigned to marker no. 1 and trace
no. 1.
CALC:MARK:FUNC:DOM:USER:STARt 1GHz
Set the start frequency of the search range to 1 GHz.
CALC:MARK:FUNC:DOM:USER:STOP 1.2GHz
Set the stop frequency of the search range to 1.2 GHz.
Manual operation:
See