GUI Reference
R&S
®
ZNC
179
User Manual 1173.9557.02 ─ 13
The limit line colors are defined in the "Define User Color Scheme" dialog ("SYSTEM >
DISPLAY > Config > Define User Color..."). You can choose between various options:
●
Display upper and lower limit lines with different colors.
●
Assign the same color to traces and associated limit lines.
●
Assign different colors to limit line segments with disabled limit check.
Note:
Display of the limit line and limit check are independent of each other: Hiding the
limit line does not switch off the limit check.
Remote command:
Limit Check
Switches the limit check of the active trace on or off.
When the limit check is switched on, a movable "PASS" or "FAIL" message is displayed
in the diagram. If the limit check fails at a measurement point, the point is marked with a
colored square. The "Limit Fail Trace" color is defined in the "Define User Color
Scheme" dialog ("Display > Display Config. > Define User Color..."). An acoustic signal
("Global Beep") and a TTL signal indicating pass or fail can be generated in addition.
The appearance of the limit fail symbols is defined in the "Define User Color Scheme"
dialog ("SYSTEM > DISPLAY > Config > Define User Color..."). You can choose between
various options:
●
Change the trace color between failed measurement points.
●
Show or hide the colored squares.
Note:
Limit check and display of the limit lines are independent of each other: With
disabled limit check, the limit line can still be displayed. If no limit lines are defined for the
active trace, the limit check can be switched on but will always PASS the trace.
Remote command:
Trace Settings